Hanna Turbidity Meters
Turbidity Meters: HI 93102
  • State of the art.
  • comply with USEPA standards
  • Logging capability

Handheld Models :
HI 93102
HI 93114
HI 93124

HI 93703

Bench Model :
LP 2000
 
Complete Water Analysis Meter
  • Good Laboratory Practice (GLP)
  • USEPA Compliance
  • Custom Calibration Points

The most important parameters needed for water analysis, especially in drinking water are now available in a portable meter. You can obtain lab performance in the field by using the HI 93102. It not only measures turbidity, but also pH, total and free chlorine, bromine, iodine, iron and cyanuric acid (CYS). All this in the palm of your hand and in seconds! Measurements are made quickly and repeatedly through a sophisticated, yet easy-to-use microprocessor. In colorimetric mode, users can select between factory preprogrammed calibration or calibrating the meter on their own, and measure either concentration or relative absorbance of the sample. Up to 25 measured samples can be stored in memory together with time and date.

HI 93102 is supplied complete with glass cuvet, batteries and instructions.

 

Available Accessories:
HI 731318 | Tissue for wiping cuvets (4 pcs)
HI 731321 | Spare measurement cuvets (4 pcs)
HI 93701-01 | Reagent kit for 100 tests (Free Cl)
HI 731327 | Rugged carrying case kit
HI 93710-01 | Reagent kit for 100 tests (pH)
HI 93711-01 | Reagent kit for 100 tests (Total Cl)
HI 93716-01 | Reagent kit for 100 tests (Br)
HI 93718-03 | Reagent kit for 300 tests (I)
HI 93722-01 | Reagent kit for 100 tests (CYS)
HI 93746-01 | Reagent kit, 100 pkt for 50 tests (Fe)
HI 93102-0 | AMCO-EPA-1 cal. Sol. @ 0 NTU (30 mL)
HI 93102-20 | AMCO-EPA-1 cal. Sol. @ 20 NTU (30 mL)
 
Specifications:  
Range Turbidity
  0.00 to 9.99 NTU & 0.0 to 50.0 NTU*
Range Bromine
  0.00 to 8.00 mg/L
Range Free Chlorine
  0.00 to 2.50 mg/L
Range Total Chlorine
  0.00 to 3.50 mg/L
Range Cyanuric Acid
  0 to 80 mg/L
Range Iodine
  0.0 to 12.5 mg/L
Range Iron
  0.00 to 1.00 mg/L
Range pH
  5.9 to 8.5 pH
Resolution Turbidity
  0.01 & 0.1 NTU
Resolution Bromine
  0.01 mg/L
Resolution Free Chlorine
  0.01 mg/L
Resolution Total Chlorine
  0.01 mg/L
Resolution Cyanuric Acid
  1 mg/L
Resolution Iodine
  0.1 mg/L
Resolution Iron
  0.01 mg/L
Resolution pH
  0.1 mg/L
Accuracy (@20°C/68°F) Turbidity
  ±0.5% NTU or ±5% of reading (whichever greater)
Accuracy (@20°C/68°F) Bromine
  ±0.8 mg/L ±3% of reading
Accuracy (@20°C/68°F) Free Chlorine
  ±0.03 mg/L ±3% of reading
Accuracy (@20°C/68°F) Total Chlorine
  ±0.03 mg/L ±3% of reading
Accuracy (@20°C/68°F) Cyanuric Acid
  ±1 mg/L ±15% of reading
Accuracy (@20°C/68°F) Iodine
  ±0.1 mg/L ±5% of reading
Accuracy (@20°C/68°F) Iron
  ±0.02 mg/L ±3% of reading
Accuracy (@20°C/68°F) pH
  ±0.1 pH
Typical EMC Deviation Turbidity
  ±2% Full Scale
Typical EMC Deviation Bromine
  ±0.01 mg/L
Typical EMC Deviation Free Chlorine
  ±0.01 mg/L
Typical EMC Deviation Total Chlorine
  ±0.01 mg/L
Typical EMC Deviation Cyanuric Acid
  ±1 mg/L
Typical EMC Deviation Iodine
  ±0.1 mg/L
Typical EMC Deviation Iron
  ±0.01 mg/L
Typical EMC Deviation pH
  ±0.1 pH
Light Source
  Pure Green LED
Light Life
  Life of the instrument
Light Detector
  Two Silicon Photocells
Battery Type / Life
  4 x 1.5V AA batteries/ 60 hours or 1,000 measurements
Environment
  32 to 122°F (0 to 50°C); RH 95%
Dimensions
  8.7 x 3.3 x 1.8" (220 x 83 x 46 mm)
Weight
  18 oz. (510 g)